To compare calculated corneal surgically induced astigmatism (SIA) by means of anterior-based keratometry (K) and total keratometry (TK) measurements made by 2 biometric devices.
Ophthalmology Department, Shaare Zedek Medical Center, Jerusalem, Israel.
Retrospective, consecutive case series.
The medical records of patients who had undergone cataract surgery through a 2.4 mm temporal clear corneal incision by a single surgeon between 3/2018 and 11/2020 were retrospectively reviewed. Patients for whom there were preoperative and postoperative keratometry measurements by 2 biometric devices, optical low-coherence reflectometer (OLCR) and swept-source optical coherence tomography (SS-OCT), were identified. Corneal SIA (mean vector value) was calculated by vector analysis for 3 groups: SS-OCT(K), SS-OCT(TK) and OLCR(K). Bivariant analyses were applied for comparisons.
One-hundred forty-seven eyes of 123 patients (73 OD and 74 OS) were enrolled in the study. The OD corneal SIA values were 0.09 D @ 136°, 0.09 D @ 141°, and 0.07 D @ 123°, for the SS-OCT(K), SS-OCT(TK), and OLCR, respectively. The respective OS corneal SIA values were 0.13 D @ 120°, 0.11 D @ 123°, and 0.08 D @ 120°. There were no significant differences between the mean vector value and variance of the corneal SIA for the right, P=.78 and P=.65, and the left, P=.75 and P=.37 eyes of the three groups.
Corneal SIA values were low (0.07 to 0.13 D) and similar for the SS-OCT and the OLCR biometric devices with standard keratometry. Total keratometry measurements yielded similar corneal SIA values compared to anterior corneal-based measurements.