To evaluate refractive error after cataract surgery using an optical biometer equipped with anterior-segment optical coherence tomography (AS-OCT).
Chukyo Eye Clinic, Nagoya, Japan.
Retrospective observational design.
In total, 150 patients with cataract (150 eyes, mean age 73.4 ± 8.2 years, men 76, women 74), who underwent measurement of parameters with the anterior-segment OCT scanners ANTERIONTM (AS-OCTB) and IOL Master 700 (OCTB) before cataract surgery, were enrolled in the study. Refractive prediction error was compared between the two devices using the SRK/T, Haigis, and Barrett UII formulas for IOL power calculation.
There were significant differences between AS-OCTB and OCTB in axial length, mean corneal refractive power, anterior chamber depth, lens thickness, and corneal diameter. In the SRK/T formula, the arithmetic means of refractive prediction errors for AS-OCTB and OCTB were -0.06 ± 0.46 D and 0.02 ± 0.42 D, respectively. In the Haigis formula, the arithmetic means of refractive prediction errors for AS-OCTB and OCTB were -0.23 ± 0.40 D and -0.08 ± 0.35 D, respectively. In the Barrett UII formula, the arithmetic means of refractive prediction errors for AS-OCTB and OCTB were -0.02 ± 0.38 D and 0.11 ± 0.36 D, respectively. AS-OCTB showed significantly larger refractive prediction error toward myopia than OCTB in all three formulas (P <0.0001).
The refractive prediction error using AS-OCTB showed a small difference from that using OCTB. While clinically comparable, the two methods could drive meaningful differences in IOL selection.
Copyright © 2021 Published by Wolters Kluwer on behalf of ASCRS and ESCRS.
About The Expert
Yukihito Kato
Takashi Kojima
Akeno Tamaoki
Kei Ichikawa
Kensei Tamura
Kazuo Ichikawa
References
PubMed