This study aims to investigate the reliability and efficacy of maximum fluorescein tear break-up time (FTBUTmax) in diagnosing dry eye disease (DED). 147 participants were enrolled in this study. Ocular symptoms were assessed by Ocular Surface Disease Index (OSDI). The fluorescein tear break-up time (FTBUT) examination, corneal fluorescein staining (CFS), and Schirmer I test were performed on both eyes. Each participant underwent 3 consecutive FTBUT tests, and five types of FTBUT values including FTBUTmax, the minimum FTBUT (FTBUTmin), the first FTBUT (FTBUT1), the average of three FTBUTs (FTBUT123) and the average of the first and second FTBUT (FTBUT12) were recorded. FTBUTmax was larger than the other FTBUT values, but no differences were found among the values of FTBUT1, FTBUT123, FTBUT12 and FTBUTmin. In the ROC analysis, FTBUTmax had the largest or the second largest area under the ROC (AUROC) in all three DED diagnostic criteria, while FTBUTmin had the least AUROC of them. ROC efficacy of FTBUTmax was significantly higher than that of FTBUT123, FTBUT12, FTBUT1 and FTBUTmin in the OSDI criteria and higher than that of FTBUT1 and FTBUTmin in Schirmer I test and CFS tests. FTBUTmax has a close correlation with OSDI, Schirmer I test and CFS, and is an effective tool for the DED diagnosis.
About The Expert
Yujie Mou
Huan Xiang
Lin Lin
Kelan Yuan
Xin Wang
Yaying Wu
Jinjin Min
Xiuming Jin
References
PubMed